Configure Alarms - ABB AV410 User Manual

Single and dual input dissolved organics monitor
Table of Contents

Advertisement

5.4 Configure Alarms

-----
CONFIG. ALARMS
-----
Config. Alarm 1
Wash
Status
Alarm
-----
Off
Off
A1: Type
Sen. B
-----
Off
Sen. A
A1: Assign
Configure Alarm 1
Alarms 2 and 3 configuration is identical to Alarm 1.
Config. Alarm 2
Alarm 1 Type
Select the type of alarm required:
– The alarm is disabled, the alarm LED is off and the relay is de-energized at
Off
all times.
Alarm
– The monitor is configured to generate an alarm in response to a specified
sensor reading.
Status
– An alarm is generated if either a power failure or a condition occurs that
causes any of the error messages in Table 9.1 (page 54) to be displayed.
Wash
– Alarm 3 is configured to control the wash sequence.
Note. The wash alarm type can be assigned only to alarm 3 and is displayed only when
the lower display shows A3: Type.
Alarm 1 Assign
Select the alarm assignment required:
Sen. A
– The monitor activates an alarm if the dissolved organics content of the
Sen. B
process fluid measured by the selected sensor exceeds or drops below the
value set in the
selected – see next page.
1 Action
set to
A1: Type
Config. Alarm 1
set to
A1: Type
A1: Failsafe
parameter, depending on the type of
Alarm 1 Set Point
or
– return to top of page.
Off
Status
– continued on next page.
Alarm
5 PROGRAMMING...