ABB AV410 User Manual page 31

Single and dual input dissolved organics monitor
Table of Contents

Advertisement

...5.9 Test Outputs and Maintenance
Hold Outputs
set to
Auto.
1
Hrs
30
Mins
Automatic Time
Maintenance
-----
Sensor A Outputs
100. 7
Org. A
99. 5
Ref. A
Peak:S=208 R=202
35. 2
mg/l C
Sensor A Reading
Sensor A Outputs
Automatic Time
If required, set a time period between 1 and 6 hours, in 30 minute increments, for which the
outputs are held when
Hold Outputs
At the default setting of
, changes in relay action and analog outputs are disabled during
Off
sensor calibration and enabled automatically at the end of the procedure.
If a time is set, changes in relay action and analog outputs are disabled during sensor
calibration but, if the calibration is not completed within the set time, the calibration is
aborted, the display returns to the Operating Page and
Continued below.
Sensor A Outputs
See Section 7.1.
FACTORY SETTINGS
Sensor A Outputs
Organics and Reference Signals
Organics total signal
Reference total signal
The peak values of the light pulse generated by the strobe lamp.
Instantaneous Reading
Internal damping function disabled – display shows instantaneous reading from sensor.
Sensor B outputs (dual input monitors only) are identical to Sensor A outputs.
Sensor B Outputs
Single input monitor – continued on next page.
Cal.Coefficients
See Section 7.1
FACTORY SETTINGS
is set to
Auto.
CAL. ABORTED
These values equate to the percentage light transmission
through the sample.
5 PROGRAMMING...