8 SENSOR MAINTENANCE...
...8.2.1 Dismantling and Cleaning – Figs. 8.1 to 8.4
View from emitter end
Remove the two screws securing the
1
mounting plate to the angle bracket.
2
Remove the four stainless steel screws
holding the flowcell halves together.
Note. Reassembly is the reverse of dismantling.
3
Insert two of the stainless steel screws into the
holes indicated and screw them in them slowly
and evenly to push the cell halves apart.
When the O-ring seal is clear of the body,
the cell halves should separate easily.
Fig. 8.3 7320 200 High Range Sensor – Splitting the Flowcell