...6.3 Installing the Sensor – Figs. 6.4 and 6.5
Dimensions in mm (in.)
191 (7.52)
110 (4.3)
Bracket fixed to
flowcell cover,
emitter side
Fig. 6.5 7320 200 High Range Sensor – Overall Dimensions and Mounting Details
Cleaner Module
373
(14.7)
Receiver Module
4 holes ø6 (0.24)
for mounting
Fix to wall
Notes.
The emitter end mounting bracket is in two parts to facilitate
emitter module removal during maintenance – see Section 7.
For maintenance purposes, allow the following minimum
clearances around the sensor:
Left (for receiver
Right (for emitter
Top (for cleaner
405 (15.9)
Standard Solution
Filler
Sample Outlet
(8 [0.3] i.d. tube)
Sample Inlet
(12 [0.47] i.d. flexible
hose connection)
Drain
(12 [0.47] i.d. flexible
hose connection)
155 (6.1)
between
centers
module
removal) – 150mm (5.9 in.)
module
removal) – 100mm (3.94 in.)
module
removal)
– 200mm (7.87 in.)
6 INSTALLATION...
Emitter Module
35