Error
Temperature fault occurred.
The LED brightness
encountered an error.
The set limit failed.
The set duty cycle failed.
The door is open.
Please close.
The device failed to reach
the required temperature.
An error occurred with the
device's turntable.
An unknown error occurred
Cause
The Form Cure L thermistor
reported an error.
The Form Cure L
encountered an error while
setting the brightness of
the LEDs.
The Form Cure L
encountered an error
while setting the current
temperature limit.
The Form Cure L
encountered an error while
setting the brightness of
the LEDs.
The door is open while the
Form Cure L is trying to begin
a cycle.
The Form Cure L did not
reach its target temperature
within 30 minutes.
The turntable was unable
to move properly due to an
issue with the motor driver or
the motor command.
Other/unspecified cause
Solution
n
•
Disconnect and reconnect the power.
•
Disconnect the power. Remove
the Form Cure L top shell and
check that the thermistor wiring is
secure. Contact
Formlabs Support
or a
certified service provider
more information.
•
Disconnect and reconnect the power.
•
Disconnect the power. Unscrew
each of the three LED modules
(accessible from within the post-cure
chamber) and check that their wiring
is secure. Contact
Formlabs Support
or a
certified service provider
more information.
•
Disconnect and reconnect the power.
•
Disconnect the power. Unscrew
each of the three LED modules
(accessible from within the post-cure
chamber) and check that their wiring
is secure. Contact
Formlabs Support
or a
certified service provider
more information.
•
Disconnect and reconnect the power.
•
Disconnect the power. Unscrew
each of the three LED modules
(accessible from within the post-cure
chamber) and check that their wiring
is secure. Contact
Formlabs Support
or a
certified service provider
more information.
Close the Form Cure L door.
Disconnect the power. Remove the
Form Cure L top shell and check that
the thermistor and heater module wiring
is secure. Contact
Formlabs Support
or a
certified service provider
more information.
Disconnect and reconnect the power.
Disconnect and reconnect the power.
Troubleshooting | 31
for
for
for
for
for