Error
The heaters do not reach
the target temperature.
The turntable does not turn.
Parts are undercured
or do not have desired
mechanical properties.
Post-cured parts have tacky
or sticky surfaces.
7.3.2
Resolving errors
If errors persist after following these steps, contact
provider
for additional guidance.
Error
The device is
currently active.
The device state is
currently invalid.
The device's duty cycle is
currently invalid.
The call to DBus failed.
The reply to the DBus
was invalid.
The fan control failed.
The timer has expired.
30
Cause
•
Abnormal display behavior
•
Environmental conditions
•
Faulty or damaged heater
•
Turntable is obstructed
•
Turntable is not fully seated
•
Turntable gear has
separated from the
turntable
•
Faulty or damaged motor
assembly
•
Expired resin
•
Part was washed but not
fully dried before post-
curing
•
Part was not washed
before post-curing
•
Part was washed but not
fully dried before post-
curing
Cause
The Form Cure L attempted
to start a post-curing cycle
while a cycle was already
in progress.
Firmware error
Firmware error
Firmware error
Firmware error
Firmware error
The Form Cure L is not
receiving temperature data.
Solution
•
Disconnect and reconnect the power.
•
Ensure that the operating
environment is in the recommended
temperature range.
•
Check that the heater fan is spinning.
•
Check and compare the internal
temperature readings from both
heater modules.
•
Ensure that no cured resin or printed
parts are blocking the turntable.
Reorient large prints as necessary.
•
Reseat the turntable on its mount.
•
Check that the gear on the bottom
of the turntable is firmly adhered to
the turntable.
•
Check that the rollers underneath the
turntable turn smoothly.
•
Replace the motor assembly.
•
Check the expected lifetime of the
resin used for the print.
•
Fully dry all solvent off of parts before
post-curing.
•
Wash liquid resin off of printed parts
before postcuring.
•
Fully dry all solvent off of parts before
post-curing.
Formlabs Support
or a
Solution
n
Disconnect and reconnect the power.
Disconnect and reconnect the power.
Disconnect and reconnect the power.
Disconnect and reconnect the power.
Disconnect and reconnect the power.
Disconnect and reconnect the power.
Disconnect and reconnect the power.
certified service