Detector Battery Activation
To ensure the safety of the battery during transportation or storage, the battery is set
to ship mode where it is locked and will not provide any voltage output.
Exit Ship Mode
Charger
AL4608
Method
2-16
Steps
1. Power on the charger.
2. Insert the battery into the charger for 3
to 5 seconds to exit ship mode.
Operation