Carestream Focus HD 35 User Manual page 10

Detector
Table of Contents

Advertisement

Input AC Power Port
Emissions Test
Electrical fast transients/burst
Surges
Line-to-line
Surges
Line-to-ground
Conducted disturbances
induced by RF fields
Voltage dips
Voltage dips
Voltage interruptions
AL4608
EMC Standard
Professional healthcare facility environment
IEC 61000-4-4
±2 kV
100 kHz repetition frequency
IEC 61000-4-5
±0.5 kV, ±1 kV
IEC 61000-4-5
±0.5 kV, ±1kV, ±2 kV
IEC 61000-4-6
3 V, 0.15 MHz–80 MHz
6 V in ISM bands between 0.15 MHz and
80 MHz
80 % AM at 1 kHz
0 % UT; 0.5 cycle
IEC 61000-4-11
At 0 º, 45 º, 90 º, 135 º, 180 º, 225 º, 270 º and
315 º
IEC 61000-4-11
0 % UT; 1 cycle
and
70 % UT; 25/30 cycles
Single phase: at 0 º
IEC 61000-4-11
0 % UT; 250/300 cycles
1-5