ANNEX
GUIDANCE AND MANUFACTURER'S
C
DECLARATION
According to: IEC 60601-1-2: 2014 (4
(Group 1, class A, for use in Hospitals) (Not LIFE-SUPPORTING)
Hatch X-Ray is tested as per applicable IEC standards, to be used under
electromagnetic environment specified below. The customer or the user of Hatch 
X-Ray should assure that it is used in such an environment.
Guidance and manufacturer's declaration – Electromagnetic
emissions – For all EQUIPMENT and SYSTEMS
Emissions test
RF emissions
IEC 60601-1-2:2014
CISPR 11:2010/
EN55011
RF emissions
IEC 60601-1-2:2014
CISPR 11:2010/
EN55011
Harmonic emissions
IEC 60601-1-2:2014
IEC 61000-3-2:2009
Voltage fluctuations 
/ flicker emissions
IEC 60601-1-2:2014
IEC 61000-3-3:2013
Table C1 : Guidance and Manufacturer's Declaration – Electromagnetic Emissions
edition)
th
Compliance
Hatch X-Ray uses RF energy only
for its internal function. Therefore, its
Group 1
RF emissions are very low and are
not likely to cause any interference in
nearby electronic equipment.
Class A
Hatch X-Ray suitable for use in all
establishments, other than domestic
establishments and those directly
Class A
connected to the public low-voltage
power supply network that supplies
buildings used for domestic purposes.
Complies
HATCH X-RAY Operating Manual
Electromagnetic environment -
guidance