Refer to Chapter "5" for details on navigating the Keypad Console
for setup and operation of Hatch X-Ray
NOTE

4.4. Power Turn-On Procedure

Turn on Hatch X-Ray by performing the following steps:
1. Place the Main Power Switch located on the bottom of the Base Unit to the ON
( I ) position. Refer "Figure 1".
Do not press any keypad keys during self test period. Any input will
be considered an error at this time.
NOTE
2. On power up, observe that the system self check function initiates and the
console displays Start-up Screen as shown by "Figure 13".
3. Immediately following a successful self test the console displays a home screen
as shown by "Figure 14".

4.5. Hatch X-Ray Operating Procedure Summary

1. Turn on Hatch X-Ray by performing the Power Turn-On Procedure provided in
section "4.4".
2.  Introduce  the  film,  plate  or  sensor  into  the  patient's  mouth  according  to  the 
chosen technique as shown by "Figure 7" and "Figure 8" (bisecting or parallel).
3. Move the tube-head beam limiter near the patient and direct it exactly towards
the tooth to be examined.
4. Arrange the tube-head with an angle suitable for the required exposure and
positioning.
5. Move as far away as the Exposure switch cable allows, in a direction opposite
to the X-ray beam emission while maintaining visual contact with the keypad
Console and the patient.
Refer to "Table 2" through "Table 7" for estimated exposure values
of minimum patient dose that can be modified per user requirement
NOTE
in Custom mode.
Create an exposure by performing the Exposure Delivery Procedure provided
in section "4.7". Make sure to refer to Chapter "5" for details on navigating the
keypad Console for setup and operation of Hatch X-Ray as necessary.