Agilent Technologies 83482A User Manual page 46

Optical/electrical plug-in module
Table of Contents

Advertisement

Calibration Overview
User Calibrations—Optical and Electrical
NOTE
The optical channel calibration accuracy is heavily dependent on the accuracy to which
you know the optical source power. For best results, measure the optical source power
with an optical power meter such as the Agilent 8153A and use precision optical con-
nectors. In addition, proper connector cleaning procedures are essential to obtaining an
accurate calibration.
To perform an O/E user-wavelength calibration
1 Press the plug-in module's front-panel optical channel SETUP key.
2 Press Calibrate, and then O/E cal.
3 Input the correct wavelength, and follow the instructions on the screen.
Figure 3-2. Plug-in calibration menu
To use an O/E user-wavelength calibration
1 Press the plug-in module's front-panel optical channel SETUP key.
2 Press Bandwidth/wavelength and then wavelength.
3 Press Usr wavelength and then Enter.
3-10

Advertisement

Table of Contents
loading

Table of Contents