Neptune Error Messages
System
Number
0.0
0.1
0.2
SYSTEM
0.3
0.4
1.0
1.1
1.2
EEPROM
1.3
1.4
1.5
Neptune 2 Error Codes
Error
Power
Distribution
reset switch
Volume
display reset
switch
These errors are not likely to occur in
the field. These are in place to ensure
System not
that unit passes testing during the
programmed
manufacturing process. If error occurs, it
Main Control
is possible circuit boards have been
Rev
Mismatch
Power Dist /
Volume
Display
Micro Rev
Mismatch
RMC
EEPROM
- EEProm failed
checksum
error
RMC
EEPROM
- Main control board not initialized
error
RMC
EEPROM
- EEProm failed
checksum
error
Small tank
calibration
board error
- Calibration board failed
Large tank
calibration
board error
Power
Distribution
- Power distribution board failed
EEPROM
read error
Possible Cause
damaged.
Trouble Shooting
N/A
- Replace Main Control Board
- Reprogram eeprom in
technician menu
- Recalibrate both canisters
- Replace Main Control Board
- Replace small or large
canister
- Calibrate float (see
Performing the Canister
Calibration on page 6-11)
- Replace power distribution
board
5-21