Fixative Solution Checking And Adjusting - J.P. SELECTA PRO-NITRO A Instruction Manual

Table of Contents

Advertisement

Available languages

Available languages

INSTRUCTION MANUAL
J
.P. SELECTA s.a.u.
CODE 80189 REV E

7.3 Fixative solution checking and adjusting

Due to the differences existing between the water used for ammonia fixa-
tive solution preparation (deionised, distilled from different assay,...) it is
necessary to check and adjust this solution to achieve a correct operation of
the equipment.
7.3.1. Material
Spectrophotometer suitable to work at 520 nm.
10 mm glass cuvette.
Flask with 0.01 ml resolution.
7.3.2 Using Boric acid solution 1% + 15 ml indicator:
(4.8 mixed, Metihyl red - Bromocresol green)
7.3.2.1
Select 520 nm in the spectrophotometer.
Adjust blank with distilled water.
Measure the fixative solution absorbance value.
The value must be approximately around 0.650 (between 0.630 and
0.670).
7.3.2.2
If the value is lower than 0.630:
Add HCl 0.1M till you get a value between 0.630 and 0.670 according to the
following formula: V = (0.650 - L) x 1515
where:
V = HCl 0.1M volume to be added per dissolution litre (ul).
L = Solution absorbance at 520 nm.
If the value is higher than 0.670:
Add NaOH 0.1M till you get a value between 0.630 and 0.670 according to
the following formula: V = (L- 0.650) x 1515
where:
V = NaOH 0.1M volume to be added per dissolution litre (ul).
L = Solution absorbance at 520 nm.
Note:
These formulas are orientative. Add a lower quantity than calculated one,
and check again the result.
The expression "ul" should be read as "microlitre".
7.3.2.3 Example:
We have a solution whose absorbance (at 520nm) is: L = 0.559
According to 7.3.2.2, we have to add:
V = (0.650 - 0.559) x 1515 ul = 137.86 ul
(of HCl 0.1 M per solution litre)
Instead of adding 137.86 ul, we add 100 ul of HCl 0.1 M.
We repeat the reading.
The new reading is: L = 0.611
We add 50 ul more and repeat the reading L = 0.640
Now we have the solution prepared to work.
Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: http://www.jpselecta.es
Checking:
Adjusting:
12/2020
Page: 52
Tel 34 937 700 877 Fax 34 937 702 362

Advertisement

Table of Contents
loading

This manual is also suitable for:

4002430

Table of Contents