Table of Contents

Advertisement

3. Cleaning the electrode cords
Clean the electrode cords by wiping them with a damp cloth.
Coating them lightly with talcum powder will reduce tangles
and help them last longer.
4. Maintenance
• Do not attempt any repairs to the device or any of its accessories.
Contact us at 1-800-363-2381 for repair information.
• The manufacturer will not be held responsible for the results
of maintenance or repairs by unauthorized persons.
• Check the device before each use for signs of wear and/or
damage. Replace worn items (electrodes, lead wires) as required.

5. To replace the battery

To ensure safe and adequate functionning of this device, it is
necessary to periodically change the battery. To do so, slide the belt
clip off, then the battery compartment cover. Remove the old
battery and insert a new one, matching the positive and negative
ends of the battery to the markings, then close the compartment.
Caution: If a battery has leaked, avoid any contact with skin, eyes
and mucous membranes. Flush any
affected skin with water and contact
a medical professional right away.
Do not attempt to recharge
the battery supplied with this
device. If you plan not to use
the device for a long time,
remove the battery and store
appropriately, away from heat.
13

THERAPY PROGRAMS

Program
Time Rate
TENS Burst
30
100Hz 200µs
Normal
30
100Hz 200µs
Mod.
30
100Hz 200µs
SD1
30
100Hz
SD2
30
100Hz
Shoulder 30
2 ~
125Hz 200µs
Lower
30
100Hz
back
Elbow
30
2 ~
125Hz 200µs
Knee
30
50Hz
EMS A
30
50Hz
S
30
50Hz
D
30
40Hz
Width
Wave description
9 pulses/burst, 2 bursts/sec.
Continuous
First 0.5 sec.: pulse width at 50%
setting. Next 0.5 sec.: pulse rate
at 50% setting and width at 100%
setting.
120 ~
Set intensity goes from 100% to
200µs
60% then back to 100% while set
pulse goes from 60% to 100% and
back to 60%, over 10-second cycles.
60 ~
Set intensity goes from 100% to
200µs
30% then back to 100% while set
pulse goes from 30% to 100% and
back to 30%, over 10-second cycles.
60 ~
12-second cycles, span: 50%.
Rate stays in the 2-10Hz range for
1/3 of the cycle time as it
modulates down to 2Hz and back
up again.
300/
1-second cycles, span: 60%.
200µs
Modulated amplitude via pulse width
0.5 sec. at 100% of set intensity and
0.5 sec. at 50%.
100/
12-second cycles, span: 50%
Rate stays in the 2-10Hz range for
1/3 of the cycle time as it
modulates down to 2Hz and back
up again.
200µs
6-second cycles (1 sec. up / 1 sec.
down) then 6 seconds rest.
200µs
6-second cycles (1 sec. up / 1 sec.
down) then 6 seconds rest.
200µs
8-second cycles(2 sec. up / 2 sec.
down) then 4 seconds rest.
200µs
11-second cycles (4 sec. up / 2 sec.
down) then 15 sec. rest (1 sec. delay).
14