The screen displays:
"Unit Overheat"
"Please Wait"
"Switch Off to"
"reset"
"menu"
The screen displays:
"Over Temp"
The screen displays:
"Heater Over Temp"
The screen displays:
"Unit Over Temp"
If your problem persists, it may be possible to rectify any issues over the
telephone. Our dedicated service and product support personnel may be able
to troubleshoot your problem remotely, thus causing minimal disruption.
Contact our After Sales Department:
Please have to hand your model and serial number together with information
on the problem prior to contacting us.
If we are unable to solve your problem over the phone, we may suggest
returning your product to the distributor where purchased or directly to
Ultrawave. Where appropriate, we operate a Return To Base (RTB) warranty
and repairs policy.
Qi Operating Manual
The Qi-Series has detected that the ultrasonic
generator on the unit has operated above the
rated
temperature
ultrasonic function. This also disables a cycle
from starting. The most likely cause is the
liquid evaporating when the level sensor is
switched off. Switch the unit off, allow the unit
to cool for a few minutes, top up the liquid
above the level sensor and switch the unit back
on.
The Qi-Series has detected that the water in
the bath is above 80ºC and has aborted the
cycle. Please allow the liquid to cool or drain
some fluid and replace with cooler liquid.
The Qi-Series has detected that the internal
water
heater
maximum temperature. Please allow the liquid
to cool or drain some fluid and replace with
cooler liquid.
The Qi-Series has overheated due to prolonged
use. Please allow a few minutes for the unit to
cool
30
and
has
has
exceeded
+44 (0) 845 330 4238
+44 (0) 29 2083 7336
service@ultrawave.co.uk
disabled
the
its
allowed
C6027205