11.1 Simulation output test

1. Finished the installation and supply the sensor with 18~30Vdc.
2. Lean a magnet close to the + sing for 2 seconds or more, there
will be switching output with corresponding LED light display.
3. Remove the magnet from the + sign, the switching output
and corresponding LED light display will return to
normal status.

11.2 M12 cable connector for dust-explosion zone

 Do not open the connector when there is an explosive dust environment.
 Please complete the M12 cable installation in the safety area first, and
then complete the on-site installation before providing supply power.
 Please use 14mm size open-end wrench to operate with the torque around
1~1.5 Nm (0.1~0.15 kgf.m)..
 Use M12 cable type in dust explosion-proof environment which approved by
NEPSI; ADOAH040VAS0005E04, ADOAH043VAS00005E04 (order code;
PC3121231415M011, PC312-1232410501)
Torque: 1.5Nm