YOKOGAWA SX42 Instruction Manual page 16

2-electrode conductivity sensors for high temperature/pressure
Hide thumbs Also See for SX42:
Table of Contents

Advertisement

16
Remarks:
1.
No revision to this drawing without prior approval of FM.
2.
Installation must be in accordance with the National Electrical Code (ANSI/NFPA 70),
ANSI/ISA-RP12.06.01, and relevant local codes.
3.
The sensor shall be installed to a certified intrinsically safe HOST with the following
maximum values: Uo= 14.4 V, Io = 116.5 mA, Po = 342.4 mW.
4.
The sensor does not provide isolation from earth. Installers shall take necessary
measures to prevent the possibility of sparking resulting from differing earth
potentials between the sensors and interconnecting equipment. This can be realized
for example by selecting interconnecting equipment which provides input-to-output
and input-to-earth isolation up to 500 V rms.
5.
Sensor Model code:
Table 3:
Model
Suffix Codes
SX42
-abcd-efgh
abcd
Cell Constant:
ef
Connection type:
g
Spare code:
h
Style code:
i
Option code:
WARNING-POTENTIONAL ELECTROSTATIC CHARGING HAZARD – SEE
6.
INSTRUCTIONS
pH sensors containing accessible plastic parts and/or external conductive parts, must
be installed and used in such a way, that dangers of ignition due to hazardous
electrostatic charges cannot occur, especially in the case that the process medium is
non-conductive.
WARNING-POTENTIONAL IGNITION HAZARD – SEE INSTRUCTIONS
Contact Conductivity sensors containing light metals, must be installed and used in
such a way that, even in the event of rare incidents, ignition sources due to impact
and friction sparks are excluded.
IM 12D07J03-01EN-P
Option Codes
/i
Four alphanumeric character (A to Z, 0 to 9 or a
hyphen)
BS ISO7/1-R1 screw thread + connector without ID-chip
BV ISO 7/1-R1 screw thread + connector with ID-chip
NS NPT screw thread + connector without ID-chip
NV NPT screw thread + connector with ID-chip
AF 2-inch 600 LBS ANSI flange + terminal block
DF DN50-PN63 EN flange + terminal block
EF DN50-PN40 EN flange + terminal block
Any one character
A
Up to ten alphanumeric characters
(A to Z, 0 to 9 or hyphen)

Advertisement

Table of Contents
loading

This manual is also suitable for:

Sc42

Table of Contents