YOKOGAWA SX42 Instruction Manual page 14

2-electrode conductivity sensors for high temperature/pressure
Hide thumbs Also See for SX42:
Table of Contents

Advertisement

14
FM-United States
Applying standards
Certificate no.*
Electrical data
Specific conditions
of use
Note 4
: Intrinsically safe, entity, for Class I, Division 1, Groups A, B, C and D;
Class I, Zone 0, AEx ia IIC, Ga (entity) for hazardous (classified) locations
when installed per control drawing D&E 2020-024-A50.
Maximum sensor input parameters:
Ui= 14.4 V; Ii= 116.5 mA; Pi= 0.3424 W; Li= 0 mH
Ci= 0 nF (BS, NS and *F type) or Ci=0.4 nF (BV and NV type).
Note 5
: Ambient temperature:
-30°C to +40°C for temperature class T6,
-30°C to +95°C for temperature class T5,
-30°C to +125°C for temperature class T4 (BV and NV type),
-30°C to +130°C for temperature class T4 (BS, NS and *F types),
-30°C to +125°C for temperature class T3 (BV and NV type),
-30°C to +165°C for temperature class T3 (BS, NS and *F types),
-30°C to +125°C for temperature class T2 (BV and NV type),
-30°C to +275°C for temperature class T2 (BS, NS and *F types).
When the sensor has been connected to non intrinsically safe equipment
which exceeds the restrictions regarding the sensor input circuits, the sensor
is not suitable anymore for intrinsically safe use.
* Certification is subject to change, due to new regulations or changes in the product itself.
When a certificate is updated the certificate is updated with a new issue number.
-
FM-United States:
FM20US0123X (effective from 03-2021)
IM 12D07J03-01EN-P
: FM Class 3600
FM Class 3610
FM Class 3810
ANSI/ISA 60079-0
ANSI/ISA 60079-11
: FM20US0123X
IS CL I, DIV 1, GP ABCD, T2...T6
CL I, ZN 0, AEx ia IIC, T2...T6 Ga
Control Drawing: D&E 2020-024-A50
: See Note 4
: See Control Drawing D&E 2020-024-A50. (Page 15-18)
Temperature classes for SX42 models are defined
T2...T6, see Note 5