6.6.3.2

Automatic Calibrations

There is a screen entitled AutoCal in the service menu which will allow you do disable the
calibrations performed by the software. There may be instances where the unit is stuck in a
continuous calibration loop and until the calibrations are disabled, it can be very difficult to
troubleshoot. Disabling the calibrations will allow you to inject signals and view traces.
Refer to the troubleshooting trees for further information on using the function of this screen.
Maintenance 6-29