SMART Triggers

Dropout: Triggers if the input signal drops out for longer than a selectable time‐
out between 1 ns and 20 s. 
Glitch: Triggers on positive or negative glitches with widths selectable from 500 
ps to 20 s or on intermittent faults (subject to bandwidth limit of oscilloscope). 
Signal or Pattern Interval: Triggers on intervals selectable from 1 ns to 20 s. 
Runt: Trigger on positive or negative runts defined by two voltage limits and 
two time limits. Select between 1 ns and 20 s. 
Slew Rate: Activates a trigger when the rising or falling edge of a pulse crosses 
two threshold levels, an upper level and a lower level. 

Automatic Setup

Autosetup: Automatically sets timebase, trigger, and sensitivity to display a 
wide range of repetitive signals. 
Vertical Find Scale: Automatically sets the vertical sensitivity and offset for the 
selected channels to display a waveform with maximum dynamic range. 

Probes

Probes: One passive probe per channel standard (WR104MXi‐A, WR104Xi‐A, 
WR204Xi‐A); optional passive and active probes are available. 
Probe System ‐ ProBus: Automatically detects and supports a wide variety of 
compatible probes 
Scale Factors: Automatically or manually selected depending on probe used 

Color Waveform Display

Type: Color 10.4‐inch flat panel TFT LCD with high resolution touch screen 
Resolution: SVGA; 800 x 600 pixels; maximum external monitor output 
resolution of 2048 x 1536 pixels 
Real Time Clock: Date, hours, minutes, and seconds displayed with waveform; 
accurate to ±50 ppm; SNTP support to synchronize to precision internet clocks 
Number of Traces: Maximum of eight traces; simultaneously displays channel, 
zoom, memory, and math traces 