AC Voltage, ACV+DCV[3] (AC Coupled)
Resolu
Range
-tion
500mV 10μV 510.00 1.00%+40 0.50%+40
5V
100μV 5.1000 1.00%+20 0.35%+15
50V
1mV
500V
10mV 510.00
750V 100mV 765.0
[1]Specifications are for sine wave inputs that are greater than 5% range.
[2]Input voltage <300Vrms.
[3]The accuracy of ACV+DCV is equal to ACV's with 10 more digits added.
* The specifications are guaranteed to an input of 750V. A beeping alarm
will go off when the input value is higher than 750V.
* Input protection of 1000V peak on all ranges.
* AC-coupled true RMS – measures the AC component of the input with up
to 400Vdc of bias on any range.
* AC Common Mode Rejection Ratio.
>60 dB at dc, 50 or 60Hz ± 0.1% (1kΩ unbalanced, slow rates)
*Input impedance 1MΩ±2% in parallel with 100pF.
AC Current, ACI+DCI[3] (AC Coupled)
Resolu
Range
-tion
500μA 10nA 510.00 1.50%+50 0.50%+40 1.50%+50 3.00%+75 0.06V max
5mA
100nA 5.1000 1.50%+40 0.50%+20 1.50%+40 3.00%+60
50mA
1μA
51.000 1.50%+40 0.50%+20 1.50%+40 3.00%+60 0.14V max
500mA 10μA 510.00 1.50%+40 0.50%+20 1.50%+40 3.00%+60[2] 1.4V max
5A
100μA 5.1000 2.0%+40 0.50%+30
10A
1mA 12.000 2.0%+40 0.50%+30
[1] The 500μA range requires an input of >35μA to meet specifications. The
5mA~10A ranges need more than 5% of full scale range to meet
specifications.
[2] Input current (5k ~ 20kHz)<330mArms.
[3]The accuracy of ACI+DCI is equal to ACI's with 10 more digits added.
Full
Scale
30-50Hz
51.000 1.00%+20 0.35%+15
x
x
Accuracy (1 year 23˚C ±5˚C) [1]
Full
Scale
30-50Hz
Accuracy (1 year 23˚C ±5˚C) [1]
50-10kHz 10K-30kHz 30K-100kHz
2.00%+60
1.00%+20
1.00%+20
0.5%+15 1.00%+20[2] 3.00%+50[2]
0.5%+15
50-2kHz
2K-5kHz
x
x
3.00%+120
3.00%+50
3.00%+50
x
x
Burden
5K-20kHz
Voltage
0.6V max
x
0.5V max
x
0.8V max