Display Readings In Test Mode - Noyes AFL OPM5-2D User Manual

Light sources, optical power meters, and related test kits
Table of Contents

Advertisement

Display Readings in Test Mode

1
-88.88
2
REF
3
WAVE ID
Single WAVE mode, or
Tri WAVE mode - second page
REF. # DESCRIPTION
Battery status indicator - displays estimated battery life remaining. The 3 bars indicate
1
the charge level. When the battery is low, the battery indicator will flash.
REF indicator - appears to indicate previously stored reference value(s).
2
On the Ref key press, the reference indicator REF 2 and reference value(s) for the currently
selected wavelength or detected WAVE ID wavelength(s) will be displayed briefly (for about 2-3
seconds).
CW mode - the reference value will be displayed in the field 5 .
Single WAVE mode - The reference value will be displayed in the field 5 .
Dual WAVE mode - The reference values will be displayed in the fields 5 and 7 .
Tri WAVE mode - The reference value will be displayed in the field 5 and 7 on the first
page and in the field 5 on the second page (replaces first page automatically in ~ 2 seconds).
WAVE ID indicator - Will be displayed if the OPM4 is receiving a wavelength ID'd light.
3
CW mode - This field displays the currently enabled wavelength.
4
Single WAVE mode - This field displays the detected WAVE ID wavelength.
Dual WAVE mode - This field displays the first detected WAVE ID wavelength.
Tri WAVE mode - This field displays the first WAVE ID wavelength and the third WAVE ID
wavelength that is shown on a second page (replaces first page automatically in ~ 2 seconds).
This field displays measured power dBm or µW or insertion loss dB at the wavelength displayed
5
in the field 4 .
If power or loss is too high or low for the OPM4 to measure, this field will display HI or LO.
CW mode - This field displays modulation frequency if detected.
6
Dual WAVE mode - This field displays the second wavelength.
Dual WAVE mode - This field displays measured power dBm or µW or insertion loss dB at the
7
wavelength displayed in the field 6 .
If power or loss is too high or low for the OPM4 to measure, this field will show HI or LO.
OPM4 Models
4
8888
nm
5
µW
dB dBm
6
8888
Hz
1
-88.88
2
REF
3
WAVE ID
-88.88
Dual WAVE mode, or
Tri WAVE mode - first page
4
8888
nm
5
µW
dB dBm
6
8888
Hz
nm
7
µW
dB dBm
9