Air Requirements; Figure 2-17: Air-Cooling (-Ac) Locations - Aerotech AGV-HP Series Hardware Manual

Hide thumbs Also See for AGV-HP Series:
Table of Contents

Advertisement

AGV-HP(O) Hardware Manual

2.5. Air Requirements

WARNING: To prevent damage to the AGV-HP(O), do not attach a water hose to an air
fitting or an air hose to a water fitting.
DANGER: High pressure air can cause severe injury.
Lock out the source and bleed off the pressure before you do service to the equip-
l
ment.
Wear eye protection.
l
An air supply is needed for AGV-HP(O) models equipped with the -AC air cooling option. A gas supply
pressure of 517 kPa to 586 kPa (75 psi to 85 psi) is required when using this feature. Gas must be
delivered through a polyurethane air hose with an outer diameter (OD) of:
AGV-HP: 6 mm OD (4 mm ID)
l
AGV-HPO: 4 mm OD (2 mm ID)
l
Unless otherwise specified, a gas flow rate of 20 SLPM (standard liters per minute) at 550 kPa (80 psi)
should be observed.
If nitrogen is used, it must be 99.99% pure and filtered to 0.25 microns. [Recommended]
l
If compressed air is used, it must be filtered to 0.25 microns, dry to 0º F dew point, and oil free.
l
The filtration requirement is necessary to prevent particles from damaging the optical surfaces of
the turning mirrors.
Figure 2-17:
Air-Cooling (-AC) Locations
www.aerotech.com
2.5. Air Requirements
45