4. Execute the Relay Test as follows:
To execute the relay test, you need the Relay Test Adapter (Agilent E5250A Option
301). Prepare the Relay Test Adapter.
a. Attach the Relay Test Adapter to the E5250A input connectors as shown in
the following figure.
b. Set switch on the Relay Test Adapter to Relay Test position.
c. Make sure no output cables are connected to any cards in the E5250A
mainframe.
d. Turn on the E5250A.
e. Press the Local/Self Test key on the E5250A front panel. The Relay Test
executes.
f. Wait until the LED in Local/Self Test key turns off.
Relay Test Adapter Connection
After executing the relay test, if the Fail LED turns on, the cause is one of the
following:
a. The Relay Test Adapter is not connected.
b. A plug-in card may be defective. Contact your nearest Agilent Technologies
Service Center.
10
Relay Test Adapter
Agilent E5252A Installation Guide, Edition 2