Blanking
Blanking sets the zero value for the test and is required on most
occasions a test is carried out. Blanking is slightly different depending
upon whether the test format is a tablet test or liquid Tubetest.
Ammonia Tablet Test
3a
Ammonia TUBETESTS
3b
Use a photometer tube with
10 mL of the water being
tested without any reagent
tablets added.
Place this in the cell holder
®
Place an unused tube in the cell
holder fitted with the Tubetest
Adapter.
Alternatively, if the water being
tested is strongly coloured
Tubetest
or cloudy, a blank can be
Adapter
prepared by adding the correct
volume of sample to a tubetest
and using it as the blank
before any Nessler reagent or
Indophenol Tablet is added.