Page 1
K43003 / K43093 Auto Time Sequence Dual Twin Foaming Characteristics Test Apparatus Operation and Instruction Manual REV B Koehler Instrument Company, Inc. 1595 Sycamore Avenue • Bohemia, New York 11716-1796 • USA Toll Free: 1-800-878-9070 (US only) • Tel: +1 631 589 3800 • Fax: +1 631 589 3815 http://www.koehlerinstrument.com •...
Page 3
NF T 60-129 This unit is tested before it leaves the factory, to ensure total functionality and compliance to the above specifications and ASTM standards. Test and inspection records are on file for verification. Jesse Kelly Application Engineer Koehler Instrument Company...
Page 5
Responsibility for Registration and Annual Reporting: Koehler will not sell directly to end users in the EU and so has no responsibility to register within each EU state and to make annual reports. Koehler declares that this responsibility is born by the importer who is the first level of the distribution chain and is subject to producer responsibility.
Page 6
WEEE Marking Since Koehler products are subject to the WEEE Directive we must display the WEEE symbol shown above in accordance with European Standard EN 50419 on the equipment. It must be indelible, at least 5mm in height, and clearly legible.
DIN 51566 provide you with the best solutions. You can depend on Koehler for a full range of accurate and 4. Federal Test Method (FTM) reliable instrumentation as well as support for your laboratory testing programs. Please do not...
K430X3 Dual Twin Foaming Characteristics Test Bath Operation and Instruction Manual Instrument Specifications 2 Safety Information and Warnings Models: K43003 Safety Considerations. this K43093 equipment may involve hazardous materials and operations. This manual does not purport to Electrical address all of the safety problems associated with...
Please note that Koehler does not supply a level with this equipment. This unit 2. Check labeling for correct orientation of requires an operating surface having a 20"...
K430X3 Dual Twin Foaming Characteristics Test Bath Operation and Instruction Manual Environmental Conditions: instrument 4 Descriptions environment must comply with the following Main Instrument Controls conditions for proper setup: No / Low Dust No direct sunlight Not near heating or AC ventilation ducts ...
Page 13
K430X3 Dual Twin Foaming Characteristics Test Bath Operation and Instruction Manual cleaning and/or servicing, please be sure to hand thumb-wheel should be set on S for disconnect unit power to avoid possible injury. seconds. 8. Temperature Probe. Detects bath 14. Timer Toggle Switch. Depress to start Soak temperature and communicates with the Timer.
77.8 IMPORTANT NOTE: digital 66.0 temperature controller for the unit comes 56.0 pre-programmed from the Koehler factory. 46.5 Please do NOT attempt to re-program the 38.3 digital temperature controller as this will 30.5 void the product warranty. If assistance is 23.0...
(Left Bath) display and comparing the value with the calibrated reference device. 90 - 150°C Koehler supplies a clear (Right Bath) silicone heat transfer fluid (part # 355-001-002) with high oxidation resistance and low Auto Tune.
Over-Temperature Protection D892 instructions and pour the test samples into the cylinders to the recommended level as The Koehler K430X3 Automatic Time Dual Twin per the aforementioned test methods. Foaming Test Bath is equipped with Over- temperature Protection (OTP) circuitry, which 2.
Glass surfaces may be cleaned using a more Over-Power Protection aggressive solvent such as acetone, if necessary. The Koehler K430X3 Automatic Time Dual Twin Foaming Test Bath is equipped with Over-power Protection circuitry, which prevents the unit from SHOCK AND BURN HAZARD: Only clean unsafe electrical conditions.
K430X3 Dual Twin Foaming Characteristics Test Bath Operation and Instruction Manual Replacement Parts Part Number Description 091-240-004 Solid State Relay, 240V Thermometer / Probe Holder 265-400-002 RTD Temperature Probe 275-103-044 Temperature Controller, 100-240V, 50/60Hz 7A 332-001-003 12 x 18 Borosilicate Bath Jar 18”...
Make sure the connector plug on the rear Warranty panel is firmly plugged in. We, at Koehler, would like to thank you for your Unit is on but bath does not heat up equipment purchase, which is protected by the following warranty.
Page 22
If the carton is discarded, it will be the purchaser’s responsibility to provide an appropriate shipping carton. Returned Goods Policy To return products for credit or replacement, please contact Koehler Customer Service with your purchase order number, packing list/invoice number, the item(s) to be returned and the reason for the return.