Summary of Contents for Stryker Precision 6209-000-000
Page 1
6 inch (width) x 6 inch Color Graphics on White Background document should not appear Booklet 20# Bond or Equivalent Description/Type: on produced Labels or IFUs. Instructions For Use Stryker Instruments Print Location: Suppliers/Services: Part Number: Rev. (269) 323-7700 (800) 253-3210 Print Center 6209-001-708 DSGN-fm-50366 rev.
Page 2
System 6 Stryker Precision ® Oscillating Tip Saw 6209-000-000 Instructions For Use ENGLISH (EN) 2015-06 6209-001-708 Rev-E www.stryker.com...
• A CAUTION highlights a product reliability issue. appropriateness of this equipment and the specific ALWAYS comply with this information to prevent technique used for each patient. Stryker, as a product damage. manufacturer, does not recommend surgical procedure or technique.
Page 4
6625-XXX-XXX series emissions or decreased electromagnetic immunity of Oscillating Tip Saw the system. DO NOT modify any system component 6720-XXX-XXX series Cartridges or accessory. 6725-XXX-XXX series • ALWAYS use a Stryker Precision Oscillating Tip Saw Cartridge with this handpiece. www.stryker.com...
Page 5
• Applied Parts - The distal end of the handpiece and the cartridge (as defined by the standards listed in the Specifications section under Product Safety Certification). Head Applied Parts Cartridge Accessory Interface FAST/STD/Safe Control Wingnut Battery Latch Index Button Battery Pack Trigger Switch www.stryker.com...
Page 6
Cartridge NOTE: The incremental marks on the length of the cartridge are for reference only. www.stryker.com...
Page 7
General warning sign blade rod ends with the drive pins on the head (see figure 1). Per European Union Directive 2012/19/EU, product must be collected separately. Do not dispose of as unsorted municipal waste. Contact local distributor for disposal information. www.stryker.com...
Page 8
4. With the drive pins engaged in the rod ends, pull the cartridge away from the handpiece until the cartridge drops down onto the surface of the head. The post will pass through the large opening in the cartridge keyhole. Figure 3 – Rotate Wingnut to Lock www.stryker.com...
Page 9
CAUTION: Before operating the handpiece, ensure the NOTES: sagittal head is locked into position. Failure to comply • This handpiece is designed to accept the Stryker may result in product damage. Large Battery Pack (REF 6215-000-000) only. NOTE: The sagittal head has eight possible cutting This battery pack can be charged in the Stryker angle positions (45-degree increments).
Page 10
• If any power loss is experienced while using a handpiece, ALWAYS replace the battery pack with a fully charged battery pack. Failure to comply may result in a drained or damaged battery pack with a shortened life. www.stryker.com...
3. Properly dispose of the single use cartridge. Battery Pack Removal Depress the battery latch and pull the battery pack out of the handpiece. Care Instructions For processing instructions and disposal/recycle information, see the care instructions manual supplied with the equipment. www.stryker.com...
6209-001-708 Rev-E Troubleshooting WARNING: DO NOT disassemble or service this equipment. NOTE: For service, contact your Stryker sales representative or call Stryker customer service. Outside the US, contact your nearest Stryker subsidiary. PROBLEM CAUSE ACTION Handpiece does not run or Battery pack is discharged.
NOTE: Specifications are approximate and may vary between devices or as a result of power supply fluctuations. Model: System 6 Stryker Precision Oscillating Tip Saw (REF 6209-000-000) 8.9 inch [226 mm] height (with large battery pack) Dimensions: 1.6 inch [41 mm] width...
Page 14
First Edition; Revisions through and including April 26, 2006 European Committee for Electrotechnical Standardization (CENELEC) EN 60601-1:2006, Medical Electrical Equipment — Part 1: General Requirements for Basic Safety and Essential Performance; IEC Corrigendum 1 (2006); IEC Corrigendum 2 (2007); CENELEC Corrigendum (2010) www.stryker.com...
Page 15
The System 6 Stryker Precision Oscillating Tip Saw (REF 6209-000-000) is intended for use in the electromagnetic environment specified below. The customer or the user of the System 6 Stryker Precision Oscillating Tip Saw (REF 6209-000-000) should assure that it is used in such an environment.
Page 16
The System 6 Stryker Precision Oscillating Tip Saw (REF 6209-000-000) is intended for use in the electromagnetic environment specified below. The customer or the user of the System 6 Stryker Precision Oscillating Tip Saw (REF 6209-000-000) should assure that it is used in such an environment.
Page 17
FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the System 6 Stryker Precision Oscillating Tip Saw (REF 6209-000-000) is used exceeds the applicable RF compliance level above, the System 6 Stryker Precision Oscillating Tip Saw (REF 6209-000-000) should be observed to verify normal operation.
Page 18
The System 6 Stryker Precision Oscillating Tip Saw (REF 6209-000-000) is intended for use in the electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the System 6 Stryker Precision Oscillating Tip Saw (REF 6209-000-000) can help prevent electromagnetic interference by maintaining...