Electromagnetic compatibility - compliance information
This device has been tested and found to comply with IEC 60601-1-2:2014 Ed 4.0 for electromagnetic
compatibility (EMC) as Class B according to CISPR 11:2009. Tests levels are listed in the accompanying
tables. Only use this Philips Avent single/double electric breast pump Advanced in combination with
the adapter supplied.
Emissions
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
Guidance and manufacturer's declaration – electromagnetic immunity
The Philips Avent single/double electric breast pump Advanced is intended for use in the
electromagnetic environment specified below. The customer or the user of the breast pump should
assure that it is used in such an environment. The essential performance functions are the accuracy of
the vacuum pressure and the backflow protection to withstand milk leakage into the motor unit. The
Philips Avent single/double electric breast pump Advanced was tested for immunity to
electromagnetic disturbances and passed using the criteria of no change in modes, vacuum accuracy
(±20 %), backflow protection and operation of the breast pump.
Immunity test
Electrostatic discharge
(ESD)
IEC 61000-4-2
Radiated RF EM fields
IEC 61000-4-3
Proximity fields from RF
wireless communications
equipment
IEC 61000-4-3
Compliance
Group 1
Class B
Complies
Class A Category
Complies
Immunity test level
±8 kV contact
±15 kV air
10 V/m
80 MHz – 2.7 GHz
80 % AM at 1 kHz
385 (27), 450 (28), 710 (9),
745 (9), 780 (9), 810 (28),
870 (28), 930 (28), 1720
(28), 1845 (28), 1970 (28),
2450 (28), 5240 (9), 5500
(9), 5785 (9) MHz (V/m)
Electromagnetic environment –
guidance
The device does not use RF energy for its
intended function. Therefore, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
The device is suitable for use in locations in
residential environments and in
establishments directly connected to a low
voltage power supply network which supplies
buildings used for domestic purposes.
Electromagnetic environment -
guidance
Floors can be a source of static electricity.
When the air is very dry (<20 % RH) the user is
advised to prefer to use the device in a room
with concrete or wooden floor (instead of
using it in a room with a synthetic floor).
The device is immune against any RF wireless
communication device such as smart phone,
Wi-Fi or Bluetooth device at a distance of
30 cm (11.8 in) or more. Avoid a closer
distance as this may result in improper
operation.