77
A vial in the oven can be changed at any time by pressing <SHIFT> + <F10>. The following security question is
displayed (Fig. 56):
Fig. 56
After confirmation via <OK> the gas supply is switched off and the needle head moves up. The vial is removed
and the new 0 – vial is inserted.
Carrying out a sample measurement / blank measurement
3.2.4
When all start conditions are met, the measurement can be started by pressing <START>. The gas supply
switches off and the needle head moves upwards (Fig. 57).
Fig. 57
The 0 – vial is removed from the oven and the sample vial is inserted (Fig. 58).
Fig. 58
After confirmation via <START>, the needle head moves down again, the gas supply is switched on and the
measurement starts (Fig. 59 and Fig. 60).