Technical Specifications

The following general specifications are applicable to all masks in this
guide.
General specifications
Environmental
conditions
Gross
dimensions
Not including
headgear
Compatibility
Technical Specifications: AcuCare F1-0
Pressure (cm H
0)
2
0 to 20
20 to 40
Therapy pressure
Dead space
information
Resistance
Technical Specifications: AcuCare F1-1
Pressure (cm H
0)
2
3 to 20
20 to 40
Operating temperature: +5°C to +40°C
Operating humidity: 15% to 95% non-condensing
Storage and transport temperature: -20°C to +60°C
Storage and transport humidity: up to 95% non-condensing
Small:
148 mm (H) x 221 mm (W) x 102 mm (D)
Medium:
158 mm (H) x 230 mm (W) x 105 mm (D)
Large:
165 mm (H) x 234 mm (W) x 107 mm (D)
For a full list of compatible devices for this mask, see the
Mask/Device Compatibility List on www.resmed.com on the Products
page under Service & Support. If you do not have internet access,
please contact your ResMed representative.
Flow (L/min)
<6
<12
0 to 40 cm H
0
2
Physical dead space (mask and elbow) using the large cushion is
342 mL.
Drop in pressure measured
(nominal)
Flow (L/min)
<6
<12
at 50 L/min: 0.4 cm H
at 100 L/min: 1.2 cm H
English
O
2
O
2
7