Attached Documents - CeraGem MASTER V4 User Manual

Cgm mb-1901
Hide thumbs Also See for MASTER V4:
Table of Contents

Advertisement

1. Explanation of mark, delivery, and storage
Classification
Form of electric shock protection
Degree of electric shock protection
Degree of protection from harmful invasion of water
Depending on the mode of operation
Medical –General Medical Equipment
E358148
AS TO ELECTRICAL SHOCK,
FIRE AND MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH
(ANSI/AAMI ES60601-1 (2005)AMD 1(2012),
CAN/CSA C22.2 NO. 60601-1:14(2014), IEC 60601-1-11 (2015)
V~ : AC (alternating current) voltage source
: On (power: connectino to the mains)
Transport and
140℉
1060hpa
storage ambient
temperature range
-68℉
500hpa
TYPE BF APPLIED PART
This way up
Use no hooks
Manufacturer
Serial number
Australian compliance mark
IP20
The product's IP classification
Medical device
The disposal of waste products / residues, etc.
CERAGEM
Condition for Transport and Storage
a) An ambient temperature range of -20℃(-68℉) to +60℃(+140℉).
b) A relative humidity range of 10% to 95%.
c) An atmospheric pressure range of 500hPa to 1060hPa
Do not disassemble and contact the customer service center when problem occurs with a fuse.
* Pull the marked area to exchange the fuse (related to service engineer)
* The equipment has to be installed at place built protective device having appropriate breaking capacity so
that prospect short-circuit current do not exceed 40 A .
44
CLASS II EQUIPMENT
TYPE BF APPLIED PART.
Ordinary
Continuous operation.
: UL Classification Mark
O : Off (power: disconnection from the mains)
Transport and
Transport and
95%
storage atmospheric
storage relative
pressure range
humidity range
10%
CLASS II EQUIPMENT
Keep dry
Fragile (handle with care)
Manufacturing date
Refer to instruction manual/booklet sign
General prohibition sign
IP21
When 'waterproof cover' cover to the product
Catalogue number
Condition for Operation
a) An ambient temperature range of +10℃(+50℉) to +40℃(+104℉).
b) A relative humidity range of 30% to 75%.
c) An atmospheric pressure range of 700hPa to 1060hPa
Capacity of a fuse : AC 250V, T10AL
Time-Lag,
10A,
Low breaking Capacity(Glass Body Fuse)

2. Attached documents

Guidance and manufacturer's declaration
- electromagnetic emission
The Model CGM MB-1901(CERAGEM CO., LTD. Automatic Thermal Massager)is intended for use
in the electromagnetic environment specified below.The customer or the user of CGM MB-1901
should assure that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment - guidance
The Model CGM MB-1901 uses RF energy
RF emissions
Group 1
only for its internal function. Therefore. Its
CISPR 11
RF emissions are very low and are not likely
to cause any interference in nearby electronic
equipment
Class B
The Model CGM MB-1901 is suitable for use
in all establishments, including domestic
Harmonics emission
establishments and those directly connected
A
IEC 61000-3-2
to the public low-voltage power supply
network that supplies buildings used for
Voltage fluctuation
Complies
domestic purposes.
IEC 61000-3-3
45