TEMPERATURE CALIBRATION
1. Ensure the soldering iron is in the soldering iron holder.
2. Turn on the power switch.
3. Press and hold the aluminum temperature adjustment knob for
about 2 seconds. The display will show "F" for Fahrenheit display
mode.
4. Press the temperature adjustment knob two more times and the
display will show "CAL", indicating you are now in Calibration
Mode.
5. Adjust the temperture knob to input the tested temperature, this will
show on the top display. Press the temperature adjustment knob to
confirm the temperature.
6. Wait for approximately 4 seconds and the program will save and
exit on it's own.
SETTING MEMORY PRESETS
1. Ensure the soldering iron is in the soldering iron holder.
2. Turn on the power switch.
3. Press the temperature adjustment knob to select the memory
preset channel (CH1/CH2/CH3).
4. While the bottom display temperature is flashing, adjust the
temperature adjustment knob to set the required temperature.
5. When the temperatures stops flashing, the setting has been saved.