Plumbing
Retrofitting the
large-surface
filter
AD water
connection for
30-1,000 kPa (UK:
100-1,000 kPa)
pressure -
pressure-
resistant
DI water ring line
160
 Do NOT shorten or otherwise damage the inlet hoses (see
diagram).
See installation diagram supplied.
If the water contains a high level of insoluble components, a large-
surface filter can be installed between the stopcock and the water
inlet hose.
The large-surface filter is available from the Miele Customer Service
Department.
This washer-disinfector is supplied as standard for a pressurised
system operating between 30-1,000 kPa (UK: 100-1,000 kPa). If the
water pressure is below 200 kPa the water intake duration will be
automatically increased.
The pressure tested hose for AD water, coded green, has a ¾ inch
threaded union for connection to the onsite stopcock for AD water.
 If the machine is not going to be connected to demineralised
(DI) water, the DI water connection has to be deactivated by the
Miele Customer Service Department. The intake hose remains on
the back of the machine.
The machine can be connected to a ring line system for DI water. For
this purpose, it must be technically adapted and the controls reset by
the Miele Customer Service Department.
Please contact the Miele Customer Service Department for further
information.