Summary of Contents for SR Instruments SR Scales SR463iR-PH Series
Page 1
Model SR463iR-PH, SR463iR-3PH Series In-Floor Platform Scale Operating and Service Manual Part No. MAN463iR-PH_200520 Page 1 of 20 by S Instruments, Inc. In-Floor Platform Scale Operating and Service Manual SInstruments, Inc. 600 Young Street, Tonawanda, NY 14150 Tel: 716-693-5977 Fax: 716-693-5854 URL: www.srscales.com Copyright 2020 SInstruments, Inc.
Model SR463iR-PH, SR463iR-3PH Series In-Floor Platform Scale Operating and Service Manual Part No. MAN463iR-PH_200520 Page 3 of 20 PACKING CHECKLIST- SR463iR-PH (Two Separate Shipments) In-Floor Platform Scale DESCRIPTION QUANTITY 1st Shipment WALL BOX ASSEMBLY 1 ea PIT FRAME ASSEMBLY 1 ea CONDUIT 1 ea...
Model SR463iR-PH, SR463iR-3PH Series In-Floor Platform Scale Operating and Service Manual Part No. MAN463iR-PH_200520 Page 4 of 20 PACKING CHECKLIST- SR463iR-3PH (Two Separate Shipments) In-Floor Platform Scale DESCRIPTION QUANTITY 1st Shipment WALL BOX ASSEMBLY 1 ea PIT FRAME ASSEMBLY 1 ea CONDUIT 1 ea...
Model SR463iR-PH, SR463iR-3PH Series In-Floor Platform Scale Operating and Service Manual Part No. MAN463iR-PH_200520 Page 5 of 20 PACKING CHECKLIST- SR463iR-4PH (Two Separate Shipments) In-Floor Platform Scale DESCRIPTION QUANTITY 1st Shipment WALL BOX ASSEMBLY 1 ea PIT FRAME ASSEMBLY 1 ea CONDUIT 1 ea...
Model SR463iR-PH, SR463iR-3PH Series In-Floor Platform Scale Operating and Service Manual Part No. MAN463iR-PH_200520 Page 6 of 20 ASSEMBLY 14 3/4" 3 3/4" STEP 1: Unpack the scale system and check parts against the PACKING CHECKLIST. If there are any missing or damaged parts, please call the WALL BOX Service Hotline at: 1-800-654-6360.
Model SR463iR-PH, SR463iR-3PH Series In-Floor Platform Scale Operating and Service Manual Part No. MAN463iR-PH_200520 Page 7 of 20 ASSEMBLY cont’d STEP 6: (Figure 4) Back fill around the perimeter of the Pit Frame Assembly with 3000PSI concrete. Trowel the concrete level with the scale frame and existing floor.
Model SR463iR-PH, SR463iR-3PH Series In-Floor Platform Scale Operating and Service Manual Part No. MAN463iR-PH_200520 Page 8 of 20 ASSEMBLY cont’d Note: This scale is designed so that the Platform Assembly is level with the Pit Frame Assembly. When tiling or carpeting the platform deck, be sure to compensate for the thickness of the material.
Model SR463iR-PH, SR463iR-3PH Series In-Floor Platform Scale Operating and Service Manual Part No. MAN463iR-PH_200520 Page 9 of 20 ASSEMBLY cont’d STEP 15: Install Front Panel Assembly (Figure 8) to the previously mounted Wall Box Assembly. Use supplied star washers and screws. Ensure that the screws are securely tightened. Figure 8: Front Panel Assembly STEP 16: Restore power to the display.
Model SR463iR-PH, SR463iR-3PH Series In-Floor Platform Scale Operating and Service Manual Part No. MAN463iR-PH_200520 Page 10 of 20 REPLACEMENT PARTS AND ACCESSORIES Part # Description SE4961-1 CABLE ASSEMBLY, 12 FT SE4961-2 CABLE ASSEMBLY, 15 FT MAN463IR-PH MANUAL HANDRAIL, SS, 34” X 34” SM7810 HANDRAIL, SS, 48”...
Model SR463iR-PH, SR463iR-3PH Series In-Floor Platform Scale Operating and Service Manual Part No. MAN463iR-PH_200520 Page 11 of 20 CLEANING and DISINFECTING CLEANING To clean the display / user interface and other scale contact areas: Use a soft cloth dampened with water and mild detergent to clean scale surfaces. ...
Model SR463iR-PH, SR463iR-3PH Series In-Floor Platform Scale Operating and Service Manual Part No. MAN463iR-PH_200520 Page 12 of 20 STORAGE and TRANSPORTATION This is a built-in scale system. To maintain proper operation of this instrumentation, storage and transport conditions should not vary outside the following conditions: Relative Humidity 0% to 85%, Ambient Temperature 14°F to 122°...
Model SR463iR-PH, SR463iR-3PH Series In-Floor Platform Scale Operating and Service Manual Part No. MAN463iR-PH_200520 Page 13 of 20 BUTTON FUNCTIONS Figure 10: Button Display WEIGH Press and hold to zero scale. Button is used to zero the system before placing the patient on the scale.
Model SR463iR-PH, SR463iR-3PH Series In-Floor Platform Scale Operating and Service Manual Part No. MAN463iR-PH_200520 Page 14 of 20 BUTTON FUNCTIONS con’t ENTER Press to save change in digits for calibration, for unit’s set-up, for time and date set-up and saving completed calibration data. UP / SELECT Press UP to adjust height up from the default for BMI calculation, to increase the scale’s “on time”, or to select a digit when setting time and date.
Page 15 of 20 THEORY OF OPERATION SR Instruments patient weighing systems are digital scales. Strain-gauge force cells convert the force of an applied weight into an analog signal. This signal is amplified by an operational amplifier and converted to a digital signal by an analog to digital converter. The digital signal is transferred to a micro-controller where it is filtered, converted to appropriate units and displayed on a liquid crystal display.
Model SR463iR-PH, SR463iR-3PH Series In-Floor Platform Scale Operating and Service Manual Part No. MAN463iR-PH_200520 Page 16 of 20 CALIBRATION NOTE: Ensure that nothing is in contact with the scale system during this procedure. Remove hands from the system when noting the displayed calibration results. APPLIED HIGH LIMIT...
Page 17
Model SR463iR-PH, SR463iR-3PH Series In-Floor Platform Scale Operating and Service Manual Part No. MAN463iR-PH_200520 Page 17 of 20 CALIBRATION cont’d SETTING CALIBRATION NOTE: Ensure that nothing is in contact with the scale system during this procedure. Remove hands from the system when noting the displayed calibration results. STEP 1: Access the Calibration Switch by breaking the Calibration Seal on the underside of the display extrusion.
Model SR463iR-PH, SR463iR-3PH Series In-Floor Platform Scale Operating and Service Manual Part No. MAN463iR-PH_200520 Page 18 of 20 TROUBLESHOOTING SYMPTOM REASON/CORRECTIVE ACTION Printer paper door is not completely closed. Close Display screen toggles “DOOR OPEN” door. There is no paper in the printer and the printer Display screen toggles “DOOR AND paper door is not completely closed.
SR Instruments, Inc.’s part, and it neither assumes nor authorizes any other person to assume for SR Instruments, Inc. any other liabilities in connection with the sale of said articles. In no event shall SR Instruments, Inc.
Model SR463iR-PH, SR463iR-3PH Series In-Floor Platform Scale Operating and Service Manual Part No. MAN463iR-PH_200520 Page 20 of 20 S Instruments, Inc. Precision & Technology in ® Perfect Balance SInstruments, Inc. 600 Young Street, Tonawanda, NY 14150 Tel: 716-693-5977 Fax: 716-693-5854 URL: www.srscales.com Copyright 2020 SInstruments, Inc.
Need help?
Do you have a question about the SR Scales SR463iR-PH Series and is the answer not in the manual?
Questions and answers