OPERATION
Making Measurements with the AC Current Probe Model MN122
Connect the black (S2) and red (S1) terminals to the 200mA range of your DMM or instrument. The
MN122 has a ratio of 1000:1. This means that for 100A
clamped, 100mA
AC
per Amp. Select the range on your DMM or instrument which best corresponds to the measured
current. If the magnitude is unknown, start with the highest range (200mA
appropriate range and resolution is reached. Clamp the probe around the conductor. Take the reading
on the meter and multiply it by 1000 to obtain the measured current (e.g., 59mA reading: 59 x 1000 =
59,000mA or 59A).
For best accuracy, avoid if possible, the proximity of other conductors which may create noise.
Tips for Making Precise Measurements
When using a current probe with a meter, it is important to select the range that provides the best
resolution. Failure to do this may result in measurement errors.
Make sure that probe jaw mating surfaces are free of dust and contamination. Contaminants cause air
gaps between the jaws, increasing the phase shift between primary and secondary. It is very critical for
power measurement.
MAINTENANCE:
Warning
For maintenance use only original factory replacement parts.
To avoid electrical shock, do not attempt to perform any servicing unless you are qualified to do so.
To avoid electrical shock and/or damage to the instrument, do not get water or other foreign agents into
the probe.
Cleaning
To ensure optimum performance, it is important to keep the probe jaw mating surfaces clean at all times.
Failure to do so may result in error in readings. To clean the probe jaws, use very fine sand paper (fine
600) to avoid scratching the jaw, then gently clean with a soft oiled cloth.
REPAIR AND CALIBRATION
You must contact our Service Center for a Customer Service Authorization number (CSA#). This will
ensure that when your instrument arrives, it will be tracked and processed promptly. Please write the CSA#
on the outside of the shipping container. If the instrument is returned for calibration, we need to know if you
want a standard calibration, or a calibration traceable to N.I.S.T. (includes calibration certificate plus
recorded calibration data).
Chauvin Arnoux
15 Faraday Drive • Dover, NH 03820 USA
Tel:
(800) 945-2362 (Ext. 360)
(603) 749-6434 (Ext. 360)
Fax:
(603) 742-2346 or (603) 749-6309
repair@aemc.com
(Or contact your authorized distributor)
Costs for repair, standard calibration, and calibration traceable to N.I.S.T. are available.
NOTE: All customers must obtain a CSA# before returning any instrument.
TECHNICAL AND SALES ASSISTANCE
If you are experiencing any technical problems, or require any assistance with the proper use or
application of this instrument, please call our technical hotline:
(800) 343-1391 • (508) 698-2115 • Fax (508) 698-2118
Chauvin Arnoux
techsupport@aemc.com
www.aemc.com
will come out of the probe leads to your DMM or instrument. The output is 1mA
®
, Inc. d.b.a. AEMC
®
®
, Inc. d.b.a. AEMC
in the conductor around which the probe is
AC
®
Instruments
Instruments
) then work down until the
AC
99-MAN 100314.v1 09/06