CEM ORACLE Operation Manual page 37

Hide thumbs Also See for ORACLE:
Table of Contents

Advertisement

24. Once the Shim process is complete, the final logs will be
displayed under Recent Results. The system will also
designate whether the results are passing by displaying a
green check or failing by displaying a red X. If a red X
appears, please contact your local CEM Service
representative.
25. Remove the Oil Standard from the magnet. Then insert
the Normalization Standard in the magnet and press OK
to start the Cal-Lock sequence. Cal-Lock should take less
than 4 minutes to complete.
Note: Before placing the Normalization standard in the
magnet, inspect the glass swirl for any bubbles trapped
between the oil. If bubbles are present, gently shake and
rotate the tube until the bubbles are fully in the top of
the glass swirl.
33

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the ORACLE and is the answer not in the manual?

Subscribe to Our Youtube Channel

Table of Contents

Save PDF