5.15 Patient Compliance Timer:
The patient compliance timer can memorize 60 sets of operation records; the
total record time is 999 hours.
After the unit is turned off, you can start to use patient compliance timer. First,
press and hold "Mode" button and turn on either amplitude knob
simultaneously to initiate patient compliance timer.
Individual treatment time:
Press "INCREASE" button (triangle button) or "DECREASE" button (inverted
triangle button) to see next record of treatment time with the number of times
or previous record of treatment time with the number of times.
Press and hold "Set" button for 3 seconds to delete the showing record. After
the record is deleted, the unit wil make a"beeping"sound.
NOTE:
1. If the treatment time is under one minute, it will not be recorded. For example, If your treatment time is 10 minutes and
30 seconds, the patient compliance timer will record 10 minutes, not 11 minutes.
2. The patient compliance timer can only record up to 999 minutes for each treatment. Therefore, if you keep using the
stimulator for over 999 minutes, it will only record 999 minutes and the record time will flash to mean the treatment time
is over 999 minutes.