Aborting Programs (Abort); Pausing Programs (Pause) - Kikusui PBZ Series User Manual

Table of Contents

Advertisement

Executing Programs
See
p. 87

Aborting Programs (ABORT)

Pausing Programs (PAUSE)

114
7
Press ENTER.
The program completes. The PBZ maintains the programmed output settings at the
time when the program completed.
The STOP key turns off, and the PBZ returns to the state where you can set the DC
signal (as if you had pressed the DC key).
Executing programs when the POWER switch is turned on
Use
in advance to specify the number of the program to execute. When the
CONFIG
POWER switch is turned on, the PBZ begins execution from the specified program.
1
Press STOP or OUTPUT.
The RUN/PAUSE key turns off, and the STOP key illuminates. If you press one of these
keys while a program is in the RUN, PAUSE, or WAIT state, the program execution
stops, and the program execution stopped screen (ABORT) appears. The PBZ
maintains the programmed output settings at the time when the program was
stopped.
If you press OUTPUT, regardless of the programmed output settings, the output is
turned off.
2
Press ENTER.
The program completes. The PBZ maintains the programmed output settings at the
time when the program completed. The STOP key turns off, and the PBZ returns to the
state where you can set the DC signal (as if you had pressed the DC key).
If you press the RUN/PAUSE key while a program is executing, the program execution pauses, and
the PAUSE screen appears. To resume program execution, press the RUN/PAUSE key again. If
you press the STOP key while program execution is paused, the program execution will be aborted.
1
Press RUN/PAUSE.
The RUN/PAUSE key blinks. The PAUSE screen appears, and the program execution
is paused. The PBZ maintains the programmed output settings at the time when the
program was paused.
PROG: The number of the program that is executing
(from 1 to 16).
LOOP: The program's current loop number.
If this number exceeds 10000, "#####" is displayed.
STEP: The step number that is being executed (1 to 1024).
Time elapsed since the step began executing
(hours : minutes : seconds).
PBZ

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Pbz20-20Pbz40-10Pbz60-6.7Pbz80-5

Table of Contents