Safety Information - Siemens ARCADIS Orbic Installation And Startup

Hide thumbs Also See for ARCADIS Orbic:
Table of Contents

Advertisement

General

Safety information

Measuring the ground wire resistance
After all work has been completed and all covers have been attached, perform the
ground wire measurement according to ARTD-002.731.17.
The ground wire resistance must not exceed 0.2 ohms
NOTICE
This document describes the individual steps required to install the 3D option in the in-
stalled base - in this case the ARCADIS Orbic 3D.
WARNING
WARNING
Siemens AG
Medical Solutions
Prior to performing work, disconnect the power plug of
the ARCADIS Orbic 3D.
Danger of possibly fatal physical injury and property damage!
Noncompliance can lead to death, physical injury or property dam-
age.
Strictly follow:
The product-specific safety information in these instruc-
tions
The general safety information in instructions
TD00-000.860.01... and
The general safety information according to ARTD part 2.
Electrical voltage!
Noncompliance can lead to death, physical injury or property dam-
age.
Removing the system covers exposes voltage-conduct-
ing parts. To avoid danger, always disconnect the sys-
tem from the line voltage before removing the covers. To
do this, unplug the power plug or switch the system
power supply off and secure the power switch against
accidental activation.
If an uninterruptible power supply (UPS) is installed in
the system, the power supply of the UPS must addition-
ally be switched free of voltage or the voltage output
plug must be unplugged.
If service in connection with electrical voltage is
required, the generator safety information in instructions
TD00-000.860.01... must be observed.
SPR2-320.814.02.01.02
01.05
Page 5 of 46