Youkey Q7 Instructions For Use Manual page 51

Pocket ultrasound system
Table of Contents

Advertisement

Tristel Duo NCU
Tristel trio wipes system
Accel® PREVention™ Wipes & RTU
CaviWipes
Sani-Cloth HB Germicidal Disposable
Wipe
Sani-Cloth Plus Germicidal Disposable
Cloth
Cautio
1
The probe must be cleaned after each use;
n:
.
2
Do not use surgical brush to clean the probe. Even soft brush may
.
damage the probe. Only soft cloth is allowed;
WARNING:
Check the following items after cleaning and disinfection:
Inspection
items
Acoustical lens Visual inspection of lens surface and connection
between lens and shell of the acoustic head
Shell
Visual inspection of shell surface
Shell
Visual inspection of shell connection
connection
Residue
Visual inspection of probe surfaces, especially
pits and narrow areas
UK
UK
CA
US
US
US
1
The product must be away from machine with high current or
.
strong magnetic.
2 Do not immerse the probe in any type of liquid or detergent.
3
Disinfection of probe with gas or heating is strictly
.
prohibited.
4
It is strictly prohibited to immerse probe socket in any
.
solution. Probe surface is waterproof but only probe head.
Please be careful while using coupling agent and cleaning
or disinfecting probe surface.
5
Users are not allowed to repair without authorization in
.
case of probe failure, Product must be sent back to the
company for repair.
Inspection methods
Chlorine dioxide
Enzymes, Chlorine Dioxide
Hydrogen Peroxide
isopropyl alcohol
Alcohol free quaternary ammonium
isopropyl alcohol
Acceptance Criteria
No damages or cracks on lens surface;
No cracks at connection between lens
and shell.
No damages or cracks on shell surface
No cracks or lack of gluing at joints of
the shell
No residual detergent or disinfectant
51