Youkey Q7 Instructions For Use Manual page 10

Pocket ultrasound system
Table of Contents

Advertisement

1.
Matters needing attention in clinical examination
technology:
2.
Do not scan fetal, newborn baby or pregnant woman for a long
time.
3.
Please follow the ALARA principle during ultrasonic
diagnosis, especially in the examination of pregnant women,
fetuses and newborns. The sound power can be reduced as much
as possible when satisfactory clinical images can be
obtained.
4.
If the APP software fails during installation or use, please
promptly provide feedback to the after-sales maintenance
personnel to avoid misdiagnosis.
CAREF
5.
The APP software will be upgraded from time to time, and new
UL:
updates will be submitted by the developer. If you want a
better experience, please pay attention to download and
upgrade.
6.
Do not use incompatible coupling agents, disinfectants,
probe protective cover, probe, puncture rack.
7.
Sterile gloves must be worn to prevent infection when using
ultrasonic probes.
8.
You must use a sterile ultrasound coupling agent. Use a
coupling agent that is in compliance with local regulatory
requirements. In addition, it is necessary to properly
manage and use the ultrasonic coupling agent to ensure that
it does not become a source of infection.
9.
Normal ultrasound does not cause burns at room temperature;
however, if the probe is placed in the same position for a
long time, the patient may be burned.
The probe cover is made of natural rubber and is used with
10.
caution for natural rubber allergy.
During the examination, the temperature of the Transvaginal
11.
transducer lens surface should be lower than 43°C.
This equipment can only be operated by qualified
medical personnel.
This manual does not introduce a clinical
examination technique. It is necessary to select the
correct inspection techniques according to the
professional training knowledge and clinical
experience.