CAUTION
1
During cleaning or disinfection, please wear antibacterial gloves to
.
prevent infection.
2
After disinfection, please use sterilizing liquid to thoroughly
.
remove the residues. Chemical residue may be harmful to human body.
3
If you can not determine the effectiveness of disinfection and
.
sterilization solution, please contact the manufacturer for product
information.
⚫ Cleaning
Please follow the instructions on the manual:
A) Wear antibacterial gloves to prevent infection.
B) Clean the probe with clean water or soapy water to remove the attached stains. Polyurethane
sponge can be used for cleaning. Avoid using a brush in order not to damage the probe.
C) After cleaning the probe, wipe off the moisture on the surface of the probe with sterile cloth
or gauze. Do not dry the probe by heating.
⚫ General disinfection
Gently wipe the probe back and forth with a medical cotton ball dipped in 75% medical grade
alcohol.
Note on disinfectant solution:As for methods of dilution, concentration and disinfection, please
refer to the instructions on concentration of disinfectant solution and disinfection methods,
provided by the manufacturer;
After cleaning, wipe the probe with sterilizing cloth or gauze, then let the alcohol evaporate
naturally on the probe surface. Do not dry the probe.
As mentioned above, market popular medical grade disinfection and cleaning products:
Solutions
Tristel Duo ULT
. water or an anti - virus solution. Immersion can cause an
electric shock or malfunction.
If the coupling agent is not completely removed after each
3
detection, it will solidify and reduce the image quality of
.
the probe.
Do not place the probe in high temperature environment (over
4
55℃) during cleaning and disinfection. High temperature may
.
cause the damage of the scanner..
Origin
UK
Active Ingredients
Chlorine dioxide
50