Parameter Adjustment About B Mode; B Mode Image Parameter - Youkey Q7 Instructions For Use Manual

Pocket ultrasound system
Table of Contents

Advertisement

Changing gray-scale image effect
Increasing frame rate of Gray-
scale Image
Change the effect of blood flow
image display
4.3 Parameter adjustment about B
mode
B mode is the system's default imaging mode. The system displays echoes in two dimensions by
assigning a brightness level based on the echo signal amplitude. To achieve the best possible
image quality, properly adjust the display brightness, gain, depth settings, viewing angle, and
exam type. Also, select an optimization setting that best matches your needs.

4.3.1 B mode image parameter

During B-mode image scanning, the image parameter area on the left side of the screen provides
real-time parameter information, which allows you to adjust the following parameters:
Adjustment method
Control panel
adjustment
SoftKey adjustment
Adjust sound power
Adjust the TGC for the corresponding target area
Adjust dynamic range
Adjust effect
Adjust frame correlation
Click [one key Optimization] to optimize one key
Reduce depth
Adjust color gain
Adjust color frequency
Adjust of deflection
Adjust afterglow
Adjust scale
Adjust speed
Adjust wall filter
Adjust color priority
Parameter term
Gain, depth, one-button optimization
Frequency, extended imaging, frame-related, dynamic range,
effect, focus, spatial compounding, image enhancement, pseudo-
color map, left and right turning, up and down, sound power, TGC,
TSI