Before Turn on the scanner, please carefully check or operate the following items:
Serial
number
Temperature, humidity and atmospheric pressure shall meet the requirements
1
for use.
The probe shall not be damaged or soiled.
2
Clean the site and the environment.
3
Probe cleaning and disinfection.
4
Press the power button to turn on the probe .
Check whether the probe is normally activated, refer to the following items to open the probe:
Serial
number
There shall be no abnormal sound, smell or overheating.
1
There is no obvious abnormal noise.
2
Check if the transducer lens surface is overheat when connect the display device to
3
the scanner.
1.
WARNI
2.
NG

3.2.2 Turn off the probe

Long press the power button to turn off the scanner, the power light goes out.
After ultrasonic examination, the ultrasonic coupling agent on the surface of the probe should be
completely erased. Otherwise, the coupling agent will be solidified on the probe head and affect
the quality of the ultrasonic image
Before and after each ultrasonic examination, the probe should be cleaned and disinfected.

3.2.3 Change the detachable transducer

Change the transducer when the scanner is in frozen or off status.
Attach the transducer to the main unit firmly.
Detach the transducer from the main unit gently..
Inspection items
Inspection items
The use of abnormal heating probes may burn patients.
If any abnormality is found, the probe is faulty.Turn off the
scanner immediately and contact the service representative.