Installing sterile components

Checking the function of the lamp

2
1
3
Leica M620 / Ref. 10 714 371 / Version -
➩ Adjust the tilt of the footswitch. Set the adjustable support as
1
required.
➩ Inspect all connections and ensure that all accessories are
seated firmly.
➩ Release the articulation brakes and set them to a light resi-
stance.
Make the joint easier to move:
➩ Release the articulation brake (1).
Make the joint more difficult to move:
➩ Tighten the articulation brake (1).
➩ Pull the retaining pin (2).
➩ Unfold the swing arm.
➩ Check the weight setting of the swing arm by raising and
lowering the microscope, correct as required; see page 10.
Caution 9
Risk of infection!
➩ Leave sufficient space around the stand to prevent
contact with non-sterile components.
Standard components:
• 1 handle, large black (tilt head)
• 2 handles, transparent
• 2 drive knobs
The handles are steam- and gas-sterilizable.
➩ Sterilize the handles and drive knobs.
➩ Attach the sterile drive knobs on binocular tube and drive
knob to block ambient illumination.
➩ Attach sterile handles to optics carrier and tilt head.
➩ Actuate the power switch of the microscope.
The main lamp will light up.
➩ Move the slide button for the quick-change lamp mount (3) to
the other side.
The auxiliary lamp will light up.
➩ Move the slide button for the quick-change lamp back to the
main lamp.
The slide button for the quick-change lamp mount must
be slid fully against its stop, otherwise the lamp will not
light.
Preparation for operation
23