Total Chlorine (Liquid Reagent) - Hanna Instruments HI97101 Instruction Manual

Ph, free chlorine, totalchlorine, cyanuric acid, iron low range, bromine & iodine photometer
Table of Contents

Advertisement

8.5. TOTAL CHLORINE (LIQUID REAGENT)

Note: Free and Total Chlorine have to be measured separately with fresh unreacted
samples, following the related procedures, if both values are desired.
REQUIRED REAGENTS
Code
Description
HI93701A-T
Total Chlorine Reagent A 3 drops
HI93701B-T
Total Chlorine Reagent B 3 drops
HI93701C-T
Total Chlorine Reagent C 1 drop
REAGENT SETS
HI93701-T
Total Chlorine Reagent - 300 tests
For other accessories see ACCESSORIES section.
MEASUREMENT PROCEDURE
• Select the
Total Chlorine (Liquid)
METHOD SELECTION section.
Note: If tutorial mode is disabled, follow the measurement procedure below. If the
tutorial mode is enabled, press Measure and follow the messages on the screen.
• Fill the cuvette with 10 mL of unreacted sample (up
to the mark). Replace the plastic stopper and the cap.
• Insert the cuvette into the holder and ensure that the
notch on the cap is positioned securely in the groove.
• Press Zero. The display will show "-0.0-" when the meter is zeroed and ready for
measurement.
Quantity
method using the procedure described in the
10 mL
33

Advertisement

Table of Contents
loading

Table of Contents