Hanna Instruments HI97101 Instruction Manual page 27

Ph, free chlorine, totalchlorine, cyanuric acid, iron low range, bromine & iodine photometer
Table of Contents

Advertisement

• Remove the cuvette.
• Add the content of one packet of
HI93701-0
Free Chlorine Reagent.
Replace the plastic stopper and the cap.
Shake gently for 20 seconds.
• Insert the cuvette into the holder and
ensure that the notch on the cap is
positioned securely in the groove.
• Press Read. The display will show a 1 minute countdown prior to the measurement.
To skip the timer, press Read twice. When the timer ends, the meter will perform the
reading. The instrument displays the results in mg/L of chlorine (Cl
INTERFERENCES
Interference may be caused by:
• Bromine, Iodine, Oxidized forms of Chromium and Manganese, Ozone
• Hardness greater than 500 mg/L CaCO
sample for approximately 2 minutes after adding the powder reagent
• Alkalinity greater than 250 mg/L CaCO
CaCO
, the color of the sample may develop only partially or rapidly fade, to remove
3
the interference neutralize the sample with diluted HCl or NaOH
, to remove the interference shake the
3
or acidity value greater than 150 mg/L
3
).
2
27