General Operations; Meter Validation: Cal Check & Calibration - Hanna Instruments HI97101 Instruction Manual

Ph, free chlorine, totalchlorine, cyanuric acid, iron low range, bromine & iodine photometer
Table of Contents

Advertisement

6. GENERAL OPERATIONS

6.1. METER VALIDATION: CAL CHECK & CALIBRATION
Validation of the
HI97101
standards. The CAL Check screen guides the user step-by-step through the validation
process and user calibration (if necessary).
WARNING: Do not use any solutions or standards other than the Hanna
®
Instruments
CAL Check Standards. For accurate validation and calibration results,
please perform these at room temperature, 18 to 25 ° C (64.5 to 77.0 ° F).
Note: CAL Check Standards will not read the specified value in measurement
mode. Protect the CAL Check cuvettes from direct sunlight by keeping them in
the original packing. Store between 5 and 30 ° C (41 to 86 ° F), do not freeze.
To perform a CAL Check:
1. Press the
CAL Check / Calibration and press Select.
The "Not Available" message or the date, time and status of the last CAL Check
will be displayed on the screen.
Note: CAL Check & Calibration is for the selected method. Free and Total Chlorine
use the same CAL Check & Calibration.
2. Press Check to start a new CAL Check. Press the
the validation process.
3. Use the functional keys to enter the certificate value of the calibration standard
found on the CAL Check Standard Certificate. Press Next to continue.
involves verifying the concentration of the certified CAL Check
key to enter menu. Use the functional keys to select
key at any time to abort
13