GE Medical Systems
REV1
SECTION 3

MESSAGES

Note:
Three types of messages may be displayed for the user on the gantry or on the Control
Console:
D
I xx:
Information message: fault message which does not inhibit the exposure.
D
E yy:
Error message: fault inhibits the exposure.
D
S zz:
Stop message: exposure is stopped due to incorrect use of equipment.
A 15-character message is displayed on the Readout located on top of the Column.
The first three characters are identification characters.
characters constitute the message.
Example:
A message displayed on the Control Console has 40 characters. The first 3
characters are identification characters. The next 28 characters comprise the
message. The 7 remaining characters are also identification characters.
Example :
When calling the Field Service of GEMS, it is essential to note each message in its
entirety.
Listed below by type are all the messages which can appear on the Column
Readout and the Control Console, with the corresponding corrective action.
3-5
SENOGRAPHE DMR+
om 5128689-2-100
The remaining 12
"I 76 FAULT"
"I76 DETECTOR FAILURE
158/011"