Temperature Max:
Temperature value for 20mA input. Valid entries -449F to
1000F (-267.2C to 537.8C).
Temperature Default:
Select default temperature condition. Valid entries -450F to
1000F (-267.8C to 537.8C).
Temperature Calibrate 4mA:
Connect 4mA source to temperature input and press 4mA button
to calibrate Min.
Temperature Calibrate 20mA:
Connect 20mA source to temperature input and press 20mA
button to calibrate Max.
Pressure Units:
Drop down menu selection:
Pressure Min:
Pressure value for 4mA input. Valid entries 0 psia to 49999 psia
(0 bar to 3447.3 bar).
Pressure Max:
Pressure value for 20mA input. Valid entries 1 psia to 50000 psia
(0.069 bar to 3447.4 bar).
Pressure Default:
Select default pressure condition. Valid entries 0 psia to 50000
psia (0 bar to 3447.4 bar).
Pressure Calibrate 4mA:
Connect 4mA source to pressure input and press 4mA button to
calibrate Min.
Pressure Calibrate 20mA:
Connect 20mA source to pressure input and press 20mA button
to calibrate Max.